RT286-05-002

Suction Dry 4-6 Surgeries

A fixed installation centraised dry aspiration plant for serving 6 surgeries. The unit MUST have the following features: Fixed installation, Serve number of surgeries: 4 – 6, Output power: Three-phase, Approximately 2kW – 3.5A, 50 Hz Speed of motor: No less than 2800 rpm Vacuum setting at approximately 160 mBar Noise level without housing: No more than 65dB(A)

Category:

Loading Viewer...

Request a Call